Automatic Tissue Processor

101 Yorco Automatic Tissue Processor

A compact and sturdy instrument designed with latest technology for complete automatic dehydration and filtration of human, animal and plant’s tissues, up to final fixing in wax. Preprogrammable provision for dehydration cycle (user selectable) is a standard feature. Glass door with lock and key arrangement provided for easy viewing also prevents unauthorised tampering of timing cycle. 

Tissues are subjected to continuous agitation by rotating S.S. Tissue basket for thorough penetration of reagents. Provision for automatic adjustable delay start is a standard feature. Imported thermostat controls S.S. Wax bath temperature and maintains it throughout the process.

Supplied complete with One S.S. Basket Rotor, One S.S. Tissue Basket, 24 Nos. S.S. Tissue Capsules, 10 Nos. One litre capacity Glass Beakers, 11 Nos. Beaker Covers, 2 Nos. S.S. Wax Baths each fitted with reputed make Thermostat, 5 Nos. 24 hours calibrated metallic timing discs, and One notch plier to program timing disc.

Suitable to work on 220 V, single phase, 50 Hz, AC supply.Capacity :– 1 Litre.

102 Yorco Automatic Tissue Processor

  The latest model incorporates an automatic safety device and locking system besides other features as in YSI–101. The safety device prevents the transfer of tissue from 10th stage to 11th stage or from 11th to 12th stage, in case the wax has not been melted properly due to power failure or any malfunctioning, thus protecting the unit and the tissues. The auto locking system checks the transfer of tissues from 12th to 1st stage to safeguard against the accidental reprocessing of the same and locks the machine. This lock can be undone by pressing of a push switch to prepare the unit for fresh use.

Supplied complete with One S.S. Basket Rotor, One S.S. Tissue Basket, 24 Nos. S.S. Tissue Capsules, 10 Nos. Glass Beakers of 1 It. capacity each, 11 Nos. Beaker Covers, 2 Nos. S.S. Wax Baths, 5 Nos. 24 hrs. calibrated Timing Discs and One notch plier to program metallic timing discs.
Suitable to work on 220 V, single phase, 50 Hz, AC supply.

103 Yorco Automatic Tissue Processor

This model is specially designed and immaculately profiled. It effectively combines aesthetic appeal with functional utility. Most effective feature of the machine is that only Basket Rotor along with the single bakelite beaker cover lifts up to change the position on to the next stage of the processing cycle. All the other beakers remain covered during each change of stage, thus keeping the loss of alcohol/reagents to a minimal level due to evaporation.

Technical Data
It progresses from its natural stage, through dehydration stages to its final embedding in wax, with 12 processing stages. Quite elaborate methods can easily be employed for preparing tissues for histology. The last two stages are thermostatically controlled wax baths. Programming the machine is very simple by cutting notches at desired timing intervals on the disc calibrated for 24 hrs. If used with single load head, 36 tissues can be processed for unequal immersion time and with double load head 72 tissues can be processed using the equal immersion time. The delay timer for 24 hours is also provided for weekend and routine delay starting.

Both the baskets keep rotating to ensure better penetration. All the beakers remain covered throughout the process. On completion of the cycle the unit automatically gets locked up to prevent double processing and the tissues remain embedded in the wax bath, until removed.

Safety Device
The built-in safety device protects the tissues by not allowing them to shift from 10th to 11th or 11th to 12th stage in case the wax is not melted. It precludes any accident due to power failure or manufacturing of the wax baths.

The equipment is supplied complete with 2 Nos. S.S. Basket Rotors, 2 Nos. S.S. Baskets, 24 Nos. S.S. Tissue capsules, 10 Nos. Borosil Glass Beakers without lip, 2 Nos. S.S. Wax Baths, 5 Nos. 24 hrs. calibrated metallic timing disc, One notch cutting plier.
Considering the requirement of different individual users, the Model is available in different Capacities.
Beaker/Wax Bath Capacity : 1/2 litre. 1 litre. 2 Liters.
Suitable to work on 220V, Single Phase 50 Hz, AC supply.
Capacity :– 1/2 litre, 1 litre, 2 liters

104 Yorco Automatic Tissue Processor

Under the guidance of eminent pathologists our R & D Division has developed yet another unique model of Tissue Processor having complete Programmer logic Controller for complete cycle. Timing & stage sequence, Wax bath temperature safety system etc. are all controlled/operated by feather touch key board with Digital Display. The entire unit looks very attractive with aesthetic appeal. The data is retained in case of power failure. Front panel comprises of the key board, digital display, control switch. The PID temperature controller have been used for controlling the wax bath temperature.

Technical Data
The 12 stage timing sequence may be programmed for each stage for duration between 1 minute to 9 hrs. 59 minute, in steps of 1 minute.

Delay Time
Delay time can be programmed upto 99 hours and will be sequenced with main control.

Wax Bath Safety Control
Wax Bath PID temperature controller gives signal if temperature is not between 50° to 65°C. Safety System will function at 10th, 11th & 12th position and will stop the main unit in stage 9 onward till the set temperature is attained in the wax bath.

Basket Rotor
Slowly rotating 2 basket rotors are also operated through the microprocessor.

The unit is supplied complete with 2 Nos. S.S. Basket Rotors, 2 Nos. S.S. Baskets, 24 Nos. S.S. Tissue Capsules, 10 Nos. S.S. beakers without lip, 2 Nos. S.S. Wax Baths (Special designed for use with PID temperature controller).
Suitable to work on 220 V, single phase, 50 Hz, AC supply.

Capacity :– 1 litre, 2 liters